Videodensitometric quantification of paravalvular regurgitation of a transcatheter aortic valve: In vitro validation

Mohammad Abdelghani, Yosuke Miyazaki, Ellen S. De Boer, Jean Paul Aben, Math Van Sloun, Todd Suchecki, Marcel Van't Veer, Osama Soliman, Yoshinobu Onuma, Robbert De Winter, Pim A.L. Tonino, Frans N. Van De Vosse, Marcel C.M. Rutten, Patrick W. Serruys

Research output: Contribution to a Journal (Peer & Non Peer)Articlepeer-review

21 Citations (Scopus)

Abstract

Aims: Videodensitometric assessment of aortography provides a periprocedural quantitation of prosthetic valve regurgitation (PVR) after transcatheter aortic valve implantation. We sought to compare the videodensitometric parameters of PVR severity to the regurgitation fraction (RF) in a controlled in vitro setting. Methods and results: In a mock circulation system, a transcatheter balloon-expandable valve inserted at the aortic valve position was gradually deformed to induce different grades of paravalvular leakage and the RF was measured with a transonic flow probe. Contrast aortography was performed and the following videodensitometric parameters were generated: left ventricle aortic regurgitation (LV-AR), LV outflow tract AR (LVOT-AR), quantitative regurgitation assessment (qRA) index, relative maximum density (relative max), and maximum upslope of the LV time-density curve. The correlation was substantial between videodensitometric parameters (LV-AR, LVOT-AR, qRA index, relative max, and maximum upslope) and RF (r2=0.96, 0.96, 0.93, 0.87, and 0.93; p<0.001 for all). LV-AR (region of interest [ROI]=entire LV) and LVOT-AR (ROI=LVOT) were not different (p=0.51) and were strongly correlated (r2=0.99) with a mean difference of 1.92% (95% limits of agreement: ±2.83). The correlations of LV-AR and LVOT-AR with RF were stronger when more than one cardiac cycle was included in the analysis (one cycle: r 2=0.85 and r 2=0.83; four cycles: r 2=0.96 and r 2=0.96, for LV-AR and LVOT-AR, respectively). Including more cycles beyond four did not improve accuracy. Conclusions: Quantitative assessment of PVR by videodensitometry of aortograms strongly correlates with the actual RF in a controlled in vitro setting. Accuracy is improved by including more than one cardiac cycle in the analysis.

Original languageEnglish
Pages (from-to)1527-1535
Number of pages9
JournalEuroIntervention
Volume13
Issue number13
DOIs
Publication statusPublished - 1 Jan 2018

Authors (Note for portal: view the doc link for the full list of authors)

  • Authors
  • Abdelghani M and Miyazaki Y and de Boer ES and Aben JP and van Sloun M and Suchecki T and van 't Veer M and Soliman O and Onuma Y and de Winter R and Tonino PAL and van de Vosse FN and Serruys PW

Fingerprint

Dive into the research topics of 'Videodensitometric quantification of paravalvular regurgitation of a transcatheter aortic valve: In vitro validation'. Together they form a unique fingerprint.

Cite this