Skip to main navigation Skip to search Skip to main content

The Early neo2 Registry: Transcatheter Aortic Valve Implantation With ACURATE neo2 in a European Population

  • Andreas Rück
  • , Won Keun Kim
  • , Mohamed Abdel-Wahab
  • , Holger Thiele
  • , Tanja K. Rudolph
  • , Alexander Wolf
  • , Jan Martin Wambach
  • , Ole De Backer
  • , Lars Sondergaard
  • , Christian Hengstenberg
  • , Mika Laine
  • , Hirokazu Miyashita
  • , Henrik Bjursten
  • , Matthias Götberg
  • , Costanza Pellegrini
  • , Stefan Toggweiler
  • , Joanna J. Wykrzykowska
  • , Osama Soliman
  • , Nawzad Saleh
  • , Christopher U. Meduri
  • Karolinska University Hospital
  • Kerckhoff Heart and Lung Centre
  • Heart Centre Leipzig
  • Universitätsklinik der Ruhr-Universität Bochum
  • Elisabeth-Krankenhaus Essen
  • University Hospital of Copenhagen - Rigshospitalet
  • Medical University Vienna
  • Helsinki University Central Hospital
  • Skane University Hospital
  • Technical University Munich
  • Lucerne Cantonal Hospital
  • University Medical Center Groningen
  • Galway University Hospital

Research output: Contribution to a Journal (Peer & Non Peer) › Article › peer-review

11 Citations (Scopus)

Abstract

BACKGROUND: ACURATE neo2 is a transcatheter aortic valve implantation system consisting of a self-expanding bioprosthetic valve with supra-annular leaflet position and featuring innovations to facilitate placement accuracy and reduce paravalvular regurgitation. METHODS AND RESULTS: The goal of the Early neo2 (Early neo2 Registry of the ACURATE neo2 TAVI Prosthesis) was to gather real-life data on safety and efficacy in a European transcatheter aortic valve implantation population treated with ACURATE neo2. Data were collected from 554 consecutive patients treated with ACURATE neo2 at 12 European sites (mean age, 82 years; 66% women; mean European System for Cardiac Operative Risk Evaluation II, 4.5%±3.8%) between September 2020 and March 2021. The composite primary end point was the occurrence of any of the following: postoperative (in-hospital) paravalvular regurgitation grade ≥2, in-hospital acute kidney injury stage 3, postoperative pacemaker implantation, 30-day death, and 30-day stroke. The primary end point occurred in 12.6% of patients. The 30-day rates for all-cause death and all stroke were 1.3% and 2.7%, respectively, and 1.5% of patients exhibited stage 3 acute kidney injury. A total of 34 patients (6.2%) received a postoperative permanent pacemaker. Per core laboratory–adjudicated echocardiographic analysis, mean postoperative aortic valve gradient was 7.6±3.3 mm Hg, and 2.8% of patients exhibited paravalvular regurgitation grade ≥2. CONCLUSIONS: In this report of postmarket use of the ACURATE neo2 valve in a real-world transcatheter aortic valve implantation population, patients exhibited favorable postoperative hemodynamics and clinical outcomes and a low rate of postoperative pacemaker implantation.

Original languageEnglish
Article numbere029464
JournalJournal of the American Heart Association
Volume12
Issue number15
DOIs
Publication statusPublished - 1 Aug 2023
Externally publishedYes

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • TAVI
  • aortic regurgitation
  • aortic stenosis

Fingerprint

Dive into the research topics of 'The Early neo2 Registry: Transcatheter Aortic Valve Implantation With ACURATE neo2 in a European Population'. Together they form a unique fingerprint.

Cite this