Left ventricular unloading and concomitant total cardiac output increase by the use of percutaneous impella recover LP 2.5 assist device during high-risk coronary intervention

Marco Valgimigli, Paul Steendijk, George Sianos, Emile Onderwater, Patrick W. Serruys

Research output: Contribution to a Journal (Peer & Non Peer)Articlepeer-review

115 Citations (Scopus)

Abstract

A number of techniques have been proposed for circulatory support during high-risk percutaneous coronary interventions (PCI), but no single approach has achieved wide acceptance so far. We report on a patient with severe left ventricular (LV) impairment who underwent a PCI with the use of a new left ventricular assist device, the Impella Recover LP 2.5 system. The effects on global cardiac output were determined by thermodilution (TD) and LV pressure-volume loops obtained by conductance catheter. The activation of the pump resulted in a rapid and sustained unloading effect of the LV. At the same time, the continuous expulsion of blood into ascending aorta throughout the cardiac cycle produced by the pump resulted in an increase of systemic overall CO, measured by the TD technique, of 1.43 L/min. The procedure was uncomplicated and the patient remained uneventful at follow-up. Our single experience gives new input for future trials to assess the effect of the Impella Recover LP 2.5 assist device on outcome in this subset of patients.

Original languageEnglish
Pages (from-to)263-267
Number of pages5
JournalCatheterization and Cardiovascular Interventions
Volume65
Issue number2
DOIs
Publication statusPublished - Jun 2005
Externally publishedYes

Keywords

  • High-risk percutaneous coronary intervention
  • Left ventricular assist device
  • Pressure-volume loop

Fingerprint

Dive into the research topics of 'Left ventricular unloading and concomitant total cardiac output increase by the use of percutaneous impella recover LP 2.5 assist device during high-risk coronary intervention'. Together they form a unique fingerprint.

Cite this