TY - JOUR
T1 - In-office assessment of blood pressure in chronic kidney disease
T2 - Usual measurement versus automated BpTRU measurement
AU - O'Shaughnessy, Michelle M.
AU - Newman, Christine A.
AU - Kinsella, Sinead M.
AU - Reddan, Donal N.
AU - Lappin, David W.
PY - 2011/6
Y1 - 2011/6
N2 - Objectives: The automated BpTRU device has been shown to improve the accuracy of in-office blood pressure assessment in hypertensive populations. We aimed to determine whether this was also true for patients with chronic kidney disease. MATERIALS AND Methods: We recorded the blood pressure of 80 hypertensive outpatients with chronic kidney disease by usual automated measurement and BpTRU automated measurement. We established whether there were any statistically significant differences in the absolute blood pressure values measured by either method and whether these differences had any impact on the assessment of blood pressure control. Results: Systolic and diastolic blood pressures were significantly lower by BpTRU measurement than by usual measurement, by 10.1±12.2 mmHg (95% confidence interval: 7.4-12.8 mmHg, P<0.001) and 2.8±10.6 mmHg (95% confidence interval: 0.4-5.1 mmHg, P=0.02), respectively. Significantly, more patients achieved their blood pressure target of 130/80 mmHg or less by BpTRU measurement than by usual measurement (72.5 vs. 48.8% for systolic blood pressure, P<0.001; 68.8 vs. 61.3% for diastolic blood pressure, P=0.02). Systolic blood pressures remained significantly lower by BpTRU measurement than by usual measurement in all predefined study subgroups (estimated glomerular filtration rate <30 vs ≥30 ml/min/1.73m2; transplant vs. nontransplant). We detected more hypotension by BpTRU measurement than by usual measurement. Conclusion: Our study suggests that the BpTRU device can negate white coat effect in patients with chronic kidney disease. The use of this device in routine clinical practice could improve the overall accuracy of in-office blood pressure assessment in this high-risk population, minimizing the potential for undertreatment and overtreatment of hypertension.
AB - Objectives: The automated BpTRU device has been shown to improve the accuracy of in-office blood pressure assessment in hypertensive populations. We aimed to determine whether this was also true for patients with chronic kidney disease. MATERIALS AND Methods: We recorded the blood pressure of 80 hypertensive outpatients with chronic kidney disease by usual automated measurement and BpTRU automated measurement. We established whether there were any statistically significant differences in the absolute blood pressure values measured by either method and whether these differences had any impact on the assessment of blood pressure control. Results: Systolic and diastolic blood pressures were significantly lower by BpTRU measurement than by usual measurement, by 10.1±12.2 mmHg (95% confidence interval: 7.4-12.8 mmHg, P<0.001) and 2.8±10.6 mmHg (95% confidence interval: 0.4-5.1 mmHg, P=0.02), respectively. Significantly, more patients achieved their blood pressure target of 130/80 mmHg or less by BpTRU measurement than by usual measurement (72.5 vs. 48.8% for systolic blood pressure, P<0.001; 68.8 vs. 61.3% for diastolic blood pressure, P=0.02). Systolic blood pressures remained significantly lower by BpTRU measurement than by usual measurement in all predefined study subgroups (estimated glomerular filtration rate <30 vs ≥30 ml/min/1.73m2; transplant vs. nontransplant). We detected more hypotension by BpTRU measurement than by usual measurement. Conclusion: Our study suggests that the BpTRU device can negate white coat effect in patients with chronic kidney disease. The use of this device in routine clinical practice could improve the overall accuracy of in-office blood pressure assessment in this high-risk population, minimizing the potential for undertreatment and overtreatment of hypertension.
KW - BpTRU
KW - blood pressure measurement
KW - chronic kidney disease
KW - hypertension
KW - renal transplant
UR - http://www.scopus.com/inward/record.url?scp=79955978920&partnerID=8YFLogxK
U2 - 10.1097/MBP.0b013e328346e0db
DO - 10.1097/MBP.0b013e328346e0db
M3 - Article
SN - 1359-5237
VL - 16
SP - 124
EP - 128
JO - Blood pressure monitoring
JF - Blood pressure monitoring
IS - 3
ER -