Impact of implantation technique and plaque morphology on strut embedment and scaffold expansion of polylactide bioresorbable scaffold – Insights from absorb Japan trial

  • Yohei Sotomi
  • , Yoshinobu Onuma
  • , Jouke Dijkstra
  • , Jeroen Eggermont
  • , Shengnan Liu
  • , Erhan Tenekecioglu
  • , Yaping Zeng
  • , Taku Asano
  • , Robbert J. de Winter
  • , Jeffrey J. Popma
  • , Ken Kozuma
  • , Kengo Tanabe
  • , Patrick W. Serruys
  • , Takeshi Kimura
  • University of Amsterdam
  • Erasmus MC
  • Leiden University Medical Center
  • Beth Israel Deaconess Medical Center
  • Teikyo University Hospital
  • Mitsui Memorial Hospital
  • Imperial College London
  • Kyoto University Hospital

Research output: Contribution to a Journal (Peer & Non Peer)Articlepeer-review

30 Citations (Scopus)

Abstract

Background: The optimal implantation technique for the bioresorbable scaffold (Absorb, Abbott Vascular) is still a matter of debate. The purpose of the present study was to evaluate the effect of implantation technique on strut embedment and scaffold expansion. Methods and Results: Strut embedment depth and scaffold expansion index assessed by optical coherence tomography (OCT) (minimum scaffold area/reference vessel area) were evaluated in the ABSORB Japan trial (OCT subgroup: 87 lesions) with respect to implantation technique using either quantitative coronary angiography (QCA) or OCT. Strut embedment was assessed at the strut level (n=667), while scaffold expansion was assessed at the lesion level (n=81). The mean embedment depth was 63±59 µm. Balloon sizing and inflation pressure had no direct effect on strut embedment. Plaque morphology affected strut embedment [nonatherosclerotic (58.9±54.3 µm), fibroatheroma (73.3±59.6 µm), fibrous plaque (59.7±51.1 µm), and fibrocalcific plaque (–3.1±61.6 µm, negative value means malapposition), P <0.001]. The balloon-artery ratio positively correlated with the expansion index. This relationship was stronger when the OCT-derived reference vessel diameter (RVD) was used as a reference for balloon selection rather than the QCA-derived one [predilatation (Pearson correlation r: QCA: 0.167 vs. OCT: 0.552), post-dilatation (QCA: 0.316 vs. OCT: 0.717)]. Conclusions: Underlying plaque morphology influenced strut embedment, whereas implantation technique had no direct effect on it. Optimal balloon sizing based on OCT-derived RVD might be recommended. However, the safety of such a strategy should be investigated in a prospective trial.

Original languageEnglish
Pages (from-to)2317-2326
Number of pages10
JournalCirculation Journal
Volume80
Issue number11
DOIs
Publication statusPublished - 2016
Externally publishedYes

Keywords

  • Bioresorbable vascular scaffold
  • Implantation technique
  • Optical coherence tomography
  • Scaffold expansion
  • Strut embedment

Fingerprint

Dive into the research topics of 'Impact of implantation technique and plaque morphology on strut embedment and scaffold expansion of polylactide bioresorbable scaffold – Insights from absorb Japan trial'. Together they form a unique fingerprint.