Abstract

Purpose: While radiofrequency catheter ablation (RFCA) creates a lesion consisting of the tissue points subjected to lethal heating, the sublethal heating (SH) undergone by the surrounding tissue can cause transient electrophysiological block. The size of the zone of heat-induced transient block (HiTB) has not been quantified to date. Our objective was to use computer modeling to provide an initial estimate. Methods and materials: We used previous experimental data together with the Arrhenius damage index (Ω) to fix the Ω values that delineate this zone: a lower limit of 0.1–0.4 and upper limit of 1.0 (lesion boundary). An RFCA computer model was used with different power-duration settings, catheter positions and electrode insertion depths, together with dispersion of the tissue’s electrical and thermal characteristics. Results: The HiTB zone extends in depth to a minimum and maximum distance of 0.5 mm and 2 mm beyond the lesion limit, respectively, while its maximum width varies with the energy delivered, extending to a minimum of 0.6 mm and a maximum of 2.5 mm beyond the lesion, reaching 3.5 mm when high energy settings are used (25 W–20s, 500 J). The dispersion of the tissue’s thermal and electrical characteristics affects the size of the HiTB zone by ±0.3 mm in depth and ±0.5 mm in maximum width. Conclusions: Our results suggest that the size of the zone of heat-induced transient block during RFCA could extend beyond the lesion limit by a maximum of 2 mm in depth and approximately 2.5 mm in width.

Original languageEnglish
Article number2163310
JournalInternational Journal of Hyperthermia
Volume40
Issue number1
DOIs
Publication statusPublished - 2023
Externally publishedYes

Keywords

  • Cardiac ablation
  • computer modeling
  • mild hyperthermia
  • moderate heating
  • transient block