Skip to main navigation Skip to search Skip to main content

Haemolysis as a first sign of thromboembolic event and acute pump thrombosis in patients with the continuous-flow left ventricular assist device heartmate II

  • S. Akin
  • , O. I. Soliman
  • , A. A. Constantinescu
  • , F. Akca
  • , O. Birim
  • , R. T. Van Domburg
  • , O. Manintveld
  • , K. Caliskan
  • Erasmus MC
  • Cardialysis BV

Research output: Contribution to a Journal (Peer & Non Peer)Articlepeer-review

28 Citations (Scopus)

Abstract

Background Despite advances in pump technology, thromboembolic events/acute pump thrombosis remain potentially life-threatening complications in patients with continuous-flow left ventricular assist devices (CF-LVAD). We sought to determine early signs of thromboembolic event/ pump thrombosis in patients with CF-LVAD, which could lead to earlier intervention. Methods We analysed all HeartMate II recipients (n = 40) in our centre between December 2006 and July 2013. Thromboembolic event/pump thrombosis was defined as a transient ischaemic attack (TIA), ischaemic cerebrovascular accident (CVA), or pump thrombosis. Results During median LVAD support of 336 days [IQR: 182–808], 8 (20%) patients developed a thromboembolic event/pump thrombosis (six TIA/CVA, two pump thromboses). At the time of the thromboembolic event/pump thrombosis, significantly higher pump power was seen compared with the no-thrombosis group (8.2 ± 3.0 vs. 6.4 ± 1.4 W, p = 0.02), as well as a trend towards a lower pulse index (4.1 ± 1.5 vs. 5.0 ± 1.0, p = 0.05) and a trend towards higher pump flow (5.7 ± 1.0 vs. 4.9 ± 1.9 L m, p = 0.06). The thrombosis group had a more than fourfold higher lactate dehydrogenase (LDH) median 1548 [IQR: 754– 2379] vs. 363 [IQR: 325–443] U/L, p = 0.0001). Bacterial (n = 4) or viral (n = 1) infection was present in 5 out of 8 patients. LDH > 735 U/L predicted thromboembolic events/ pump thrombosis with a positive predictive value of 88%. Conclusions In patients with a CF-LVAD (HeartMate II), thromboembolic events and/or pump thrombosis are associated with symptoms and signs of acute haemolysis as manifested by a high LDH, elevated pump power and decreased pulse index, especially in the context of an infection.

Original languageEnglish
Pages (from-to)134-142
Number of pages9
JournalNetherlands Heart Journal
Volume24
Issue number2
DOIs
Publication statusPublished - Feb 2016
Externally publishedYes

Keywords

  • Haemolysis
  • HeartMate II
  • Left ventricular assist device (LVAD)
  • Pump thrombosis
  • Thromboembolic event

Fingerprint

Dive into the research topics of 'Haemolysis as a first sign of thromboembolic event and acute pump thrombosis in patients with the continuous-flow left ventricular assist device heartmate II'. Together they form a unique fingerprint.

Cite this