TY - JOUR
T1 - Digital health care solution for proactive heart failure management with the Cordella Heart Failure System
T2 - results of the SIRONA first-in-human study
AU - Mullens, Wilfried
AU - Sharif, Faisal
AU - Dupont, Matthias
AU - Rothman, Alexander M.K.
AU - Wijns, William
N1 - Publisher Copyright:
© 2020 The Authors. European Journal of Heart Failure published by John Wiley & Sons Ltd on behalf of European Society of Cardiology.
PY - 2020/10/1
Y1 - 2020/10/1
N2 - Aims: Incorporation of remote monitoring of pulmonary artery pressure and vital signs has been demonstrated to reduce heart failure (HF) hospitalization and all-cause mortality in selected symptomatic HF patients. The aim of this study is to investigate the safety and accuracy of the new CordellaTM Pulmonary Artery Pressure Sensor (Endotronix, Inc., Chicago, IL, USA) and the usability of the comprehensive CordellaTM Heart Failure System (CHFS). Methods and results: Multicentre, open-label, first-in-human, feasibility study to evaluate the CHFS and the safety and accuracy of the Cordella™ Pulmonary Artery Pressure Sensor in 15 patients with New York Heart Association class III HF. All patients were successfully implanted with the Cordella Pulmonary Artery Pressure Sensor, without sensor failure. No device system-related complications, defined as invasive treatment, device explant or death, occurred. The primary efficacy endpoint of a mean pulmonary artery pressure at 90 days was met in all but one patients with a cohort difference of 2.7 mmHg (Cordella Sensor 22.5 ± 11.8 mmHg, Swan–Ganz catheter 25.2 ± 8.5 mmHg). One patient did not go through the 90-day right heart catheterization for safety reasons. Patient adherence to daily measurement, transmission of vital signs and pulmonary artery pressure sensor readings were recorded 99% of the time. Conclusion: The initial experience of the CHFS incorporating comprehensive vital signs and pulmonary artery pressure monitoring enables safe and accurate monitoring of HF status.
AB - Aims: Incorporation of remote monitoring of pulmonary artery pressure and vital signs has been demonstrated to reduce heart failure (HF) hospitalization and all-cause mortality in selected symptomatic HF patients. The aim of this study is to investigate the safety and accuracy of the new CordellaTM Pulmonary Artery Pressure Sensor (Endotronix, Inc., Chicago, IL, USA) and the usability of the comprehensive CordellaTM Heart Failure System (CHFS). Methods and results: Multicentre, open-label, first-in-human, feasibility study to evaluate the CHFS and the safety and accuracy of the Cordella™ Pulmonary Artery Pressure Sensor in 15 patients with New York Heart Association class III HF. All patients were successfully implanted with the Cordella Pulmonary Artery Pressure Sensor, without sensor failure. No device system-related complications, defined as invasive treatment, device explant or death, occurred. The primary efficacy endpoint of a mean pulmonary artery pressure at 90 days was met in all but one patients with a cohort difference of 2.7 mmHg (Cordella Sensor 22.5 ± 11.8 mmHg, Swan–Ganz catheter 25.2 ± 8.5 mmHg). One patient did not go through the 90-day right heart catheterization for safety reasons. Patient adherence to daily measurement, transmission of vital signs and pulmonary artery pressure sensor readings were recorded 99% of the time. Conclusion: The initial experience of the CHFS incorporating comprehensive vital signs and pulmonary artery pressure monitoring enables safe and accurate monitoring of HF status.
KW - Heart failure
KW - Pulmonary artery pressure sensor
UR - http://www.scopus.com/inward/record.url?scp=85085621113&partnerID=8YFLogxK
U2 - 10.1002/ejhf.1870
DO - 10.1002/ejhf.1870
M3 - Article
C2 - 32476191
AN - SCOPUS:85085621113
SN - 1388-9842
VL - 22
SP - 1912
EP - 1919
JO - European Journal of Heart Failure
JF - European Journal of Heart Failure
IS - 10
ER -