Comparison of valve performance of the mechanically expanding Lotus and the balloon-expanded SAPIEN3 transcatheter heart valves: An observational study with independent core laboratory analysis

  • Osama I.I. Soliman
  • , Nahid El Faquir
  • , Ben Ren
  • , Ernest Spitzer
  • , Lennart Van Gils
  • , Hans Jonker
  • , Marcel L. Geleijnse
  • , Gerrit Anne Van Es
  • , Jan G. Tijssen
  • , Nicolas M. Van Mieghem
  • , Peter P.T. De Jaegere
  • Erasmus MC
  • Cardialysis BV

Research output: Contribution to a Journal (Peer & Non Peer) › Article › peer-review

11 Citations (Scopus)

Abstract

The Lotus and SAPIEN3 are second-generation transcatheter heart valves, which are designed to minimize paravalvular aortic regurgitation (PAR) after transcatheter aortic valve replacement (TAVR). We sought to compare both devices for valve performance and with emphasis on PAR by independent core laboratory analysis. Methods and results A total of 162 (79 Lotus and 83 SAPIEN3) consecutive patients (51% female, 8068 years, Logistic EuroSCORE 14.869.4%) who underwent TAVR because of aortic stenosis were included. Patients with aortic valve-in-valve treatment were excluded. Pre-discharge echocardiograms were analysed by an independent core laboratory using the Valve Academic Research Consortium 2 criteria. There were no differences in baseline and procedural characteristics, except for a larger aortic annulus and sizing indices in SAPIEN3-Treated patients and frequency of postdilatation (0% in Lotus and 13.1% in SAPIEN3). Both valves have similar mean residual gradient, indexed effective orifice area and Doppler velocity index when adjusted to valve size. The frequency of mild (13.9% vs. 31.3%) and at least moderate (1.3% vs. 3.6%) PAR was less after Lotus than after SAPIEN3 implantation (P%0.02). Multi-slice computed tomography-based annulus and left ventricular outflow tract diameters, calcification and percentage of oversizing were not different between those with or without mild or more PAR. On multivariate analysis, the use of Lotus valve was associated with less (odds ratio OR, 0.41, P%0.03) occurrence of PAR.

Original languageEnglish
Pages (from-to)157-167
Number of pages11
JournalEuropean Heart Journal Cardiovascular Imaging
Volume19
Issue number2
DOIs
Publication statusPublished - 1 Feb 2018
Externally publishedYes

Keywords

  • Aortic
  • Echocardiography
  • Haemodynamic
  • Outcome
  • Regurgitation
  • Stenosis
  • TAVR
  • Valve