Comparison of in vivo eccentricity and symmetry indices between metallic stents and bioresorbable vascular scaffolds: Insights from the ABSORB and SPIRIT trials

  • Salvatore Brugaletta
  • , Josep Gomez-Lara
  • , Roberto Diletti
  • , Vasim Farooq
  • , Robert Jan Van Geuns
  • , Bernard De Bruyne
  • , Dariusz Dudek
  • , Hector M. Garcia-Garcia
  • , John A. Ormiston
  • , Patrick W. Serruys

Research output: Contribution to a Journal (Peer & Non Peer)Articlepeer-review

54 Citations (Scopus)

Abstract

Objective: To compare the geometrical parameters of a bioresorbable vascular scaffold (BVS) with a standard metallic stent. Background: The introduction of polymeric bioresorbable materials in the design of novel coronary scaffolds may affect some geometrical parameters, such as eccentricity and symmetry indices, previously introduced as IVUS criteria for optimal metallic stent deployment. Methods: From ABSORB Cohort A, ABSORB Cohort B, SPIRIT I, and SPIRIT II, all patients implanted with BVS 1.0, BVS 1.1, or XIENCE V, respectively and intravascular ultrasound analyses post-implantation were selected. The eccentricity index was calculated frame by frame and expressed as an average per device (minimum diameter/maximum diameter). The symmetry index of the device was reported as ([maximum diameter - minimum diameter]/maximum diameter).Six months major adverse cardiac events (MACE) were analyzed. Results: A total of 242 patients were selected (BVS 1.0: n = 28, BVS 1.1: n = 94, XIENCE V: n = 120). The BVS exhibited a significantly lower eccentricity index (BVS 1.0: 0.83 ± 0.09; BVS 1.1: 0.85 ± 0.08; XIENCE V: 0.90 ± 0.06; P < 0.01) and a significantly higher symmetry index (BVS 1.0: 0.30 ± 0.07; BVS 1.1: 0.31 ± 0.06, XIENCE V 0.26 ± 0.07; P < 0.01) as compared to the XIENCE V. An inverse correlation was found between the symmetry and eccentricity indices for both (BVS r = -0.69, P < 0.01; XIENCE V r = -0.61, P < 0.01). No differences in MACE were detected between the groups according to their geometrical parameters. Conclusions: The introduction of a new polymeric material in the design of BVS resulted in a lower eccentricity index and a higher symmetry index as compared to metallic stents, without detectable impact in MACE, at 6 months.

Original languageEnglish
Pages (from-to)219-228
Number of pages10
JournalCatheterization and Cardiovascular Interventions
Volume79
Issue number2
DOIs
Publication statusPublished - 1 Feb 2012
Externally publishedYes

Keywords

  • BVS
  • eccentricity
  • IVUS
  • symmetry
  • XIENCE V

Fingerprint

Dive into the research topics of 'Comparison of in vivo eccentricity and symmetry indices between metallic stents and bioresorbable vascular scaffolds: Insights from the ABSORB and SPIRIT trials'. Together they form a unique fingerprint.

Cite this