Comparison of BACTEC MGIT 960 and BACTEC 460 for culture of Mycobacteria

Martin Cormican, Geraldine Corbett-Feeney

Research output: Contribution to a Journal (Peer & Non Peer)Articlepeer-review

16 Citations (Scopus)

Abstract

We have compared the BACTEC 460 system with the BACTEC MGIT 960 system for culture of mycobacteria from 1800 routine clinical specimens. Rate of isolation of M. tuberculosis and time to detection of positive culture was comparable for both systems (BACTEC 460, 35 isolates, BACTEC MGIT 960, 34 isolates). Contamination of cultures was more common with the BACTEC MGIT 960 system. With intensification of the decontamination process an acceptable contamination rate was achieved in the BACTEC MGIT 960 system but time to detection of positive culture was increased by 1 to 2 days. (C) 2000 Elsevier Science Inc. All rights reserved.
Original languageEnglish (Ireland)
Pages (from-to)123-126
Number of pages4
JournalDiagnostic Microbiology And Infectious Disease
Volume38
Issue number2
Publication statusPublished - 1 Oct 2000

Keywords

  • Bactec 460
  • Bactec MGIT
  • Culture
  • Mycobacterium

Authors (Note for portal: view the doc link for the full list of authors)

  • Authors
  • Whyte, T,Cormican, M,Hanahoe, B,Doran, G,Collins, T,Corbett-Feeney, G

Fingerprint

Dive into the research topics of 'Comparison of BACTEC MGIT 960 and BACTEC 460 for culture of Mycobacteria'. Together they form a unique fingerprint.