Comparison between the first and second generation bioresorbable vascular scaffolds: A six month virtual histology study

  • Salvatore Brugaletta
  • , Hector M. Garcia-Garcia
  • , Roberto Diletti
  • , Josep Gomez-Lara
  • , Scot Garg
  • , Yoshinobu Onuma
  • , Eun Seok Shin
  • , Robert Jan Van Geuns
  • , Bernard De Bruyne
  • , Dariusz Dudek
  • , Leif Thuesen
  • , Bernard Chevalier
  • , Dougal McClean
  • , Stephan Windecker
  • , Robert Whitbourn
  • , Cecile Dorange
  • , Susan Veldhof
  • , Richard Rapoza
  • , Krishnankutty Sudhir
  • , Nico Bruining
  • John A. Ormiston, Patrick W. Serruys
  • Erasmus MC
  • Cardiovascular Center Aalst
  • Uniwersytet Jagielloński
  • Aarhus University Hospital
  • Institut Hospitalier Jacques Cartier
  • Christchurch Hospital New Zealand
  • University Hospital of Psychiatry
  • St Vincent's Hospital
  • Abbott Vascular

Research output: Contribution to a Journal (Peer & Non Peer) › Article › peer-review

15 Citations (Scopus)

Abstract

Aims: To compare the intravascular ultrasound virtual histology (IVUS-VH) appearance of the polymeric struts of the first (Revision 1.0) and the second (Revision 1.1) generation bioresorbable vascular scaffold (BVS). Methods and results: IVUS-VH misrepresents polymeric struts as dense calcium (DC) and necrotic core (NC) so that their presence and disappearance could be used as potential artifactual surrogate of bioresorption. DC and NC were assessed in both revisions of the BVS by analysing IVUS-VH from all patients in the ABSORB cohort A (Revision 1.0) and cohort B (Revision 1.1) study who had an IVUS-VH post-treatment and at 6-month follow-up. Post-treatment and 6-month follow-up IVUS-VH results, available in 60 patients (BVS 1.0 n=28; BVS 1.1 n=32), indicated an insignificant rise in DC+NC area compared to baseline with Revision 1.1 (0.10±0.46mm2, p=0.2), whilst a significant reduction was seen with Revision 1.0 (-0.57±1.3 mm2, p=0.02). A significant correlation has been found between the change in the DC+NC area and the change in external elastic membrane area (y=0.68x-0.1; r=0.58, p=0.03). Conclusions: Based on 6-months IVUS-VH analysis, the BVS 1.1 appears to have a different backscattering signal compared to the BVS 1.0, which may reflect differences in the speed of chemical and structural alteration.

Original languageEnglish
Pages (from-to)1110-1116
Number of pages7
JournalEuroIntervention
Volume6
Issue number9
DOIs
Publication statusPublished - Apr 2011
Externally publishedYes

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • Bioresorbable vascular scaffold
  • Intravascular ultrasound virtual histology

Fingerprint

Dive into the research topics of 'Comparison between the first and second generation bioresorbable vascular scaffolds: A six month virtual histology study'. Together they form a unique fingerprint.

Cite this