Comparative analysis of deformable mirrors for ocular adaptive optics

  • Eugenie Dalimier
  • , Chris Dainty

Research output: Contribution to a Journal (Peer & Non Peer)Articlepeer-review

136 Citations (Scopus)

Abstract

We have evaluated the ability of three commercially available deformable mirrors to compensate the aberrations of the eye using a model for aberrations developed by Thibos, Bradley and Hong. The mirrors evaluated were a 37 actuator membrane mirror and 19 actuator piezo mirror (OKO Technologies) and a 35 actuator bimorph mirror (AOptix Inc). For each mirror, Zernike polynomials and typical ocular aberrated wavefronts were fitted with the mirror modes measured using a Twyman-Green interferometer. The bimorph mirror showed the lowest root mean square error, although the 19 actuator piezo device showed promise if extended to more actuators. The methodology can be used to evaluate new deformable mirrors as they become available.

Original languageEnglish
Pages (from-to)4275-4285
Number of pages11
JournalOptics Express
Volume13
Issue number11
DOIs
Publication statusPublished - 30 May 2005

Fingerprint

Dive into the research topics of 'Comparative analysis of deformable mirrors for ocular adaptive optics'. Together they form a unique fingerprint.