Autohesion of plasma treated semi-crystalline PEEK: Comparative study of argon, nitrogen and oxygen treatments

Shengnan Zhang, Firas Awaja, Natalie James, David R. McKenzie, Andrew J. Ruys

Research output: Contribution to a Journal (Peer & Non Peer)Articlepeer-review

87 Citations (Scopus)

Abstract

Semi-crystalline polyetheretherketone (PEEK) is of interest for providing hermetic sealing of implantable medical devices. Self bonding or autohesion is achieved by mild temperature and pressure treatment and is potentially useful to form joints in the encapsulation of active medical implants. The surfaces of PEEK films were treated in a radio-frequency (RF) plasma containing one of the gases Ar, N2 and O2, to achieve surface activation. The bond strength developed over the interface of PEEK films was evaluated by lap-shear testing. The effects of plasma conditions on the surface morphology, composition, and properties were determined using the profilometer, contact angle measurements, X-ray photoelectron spectrometry (XPS) and scanning electron microscopy (SEM). Results obtained show that plasma treatment of the PEEK films enhances their bonding strength, with the Ar treated films exhibiting the highest bond strength and nitrogen the lowest. Bond strength was shown to correlate positively with total oxygen content, with C-O group concentration and with the polar component of surface energy. Bond strength correlated negatively with CO group concentration.

Original languageEnglish
Pages (from-to)88-95
Number of pages8
JournalColloids and Surfaces A: Physicochemical and Engineering Aspects
Volume374
Issue number1-3
DOIs
Publication statusPublished - 20 Jan 2011
Externally publishedYes

Keywords

  • Autohesion
  • Medical encapsulation
  • PEEK
  • Plasma bonding
  • XPS

Fingerprint

Dive into the research topics of 'Autohesion of plasma treated semi-crystalline PEEK: Comparative study of argon, nitrogen and oxygen treatments'. Together they form a unique fingerprint.

Cite this