Assessing the accuracy of measuring leg length discrepancy and genu varum/valgum using a markerless motion analysis system

Robert J. Hurley, Martin S. Davey, Micheal Newell, Aiden Devitt

Research output: Contribution to a Journal (Peer & Non Peer)Articlepeer-review

8 Citations (Scopus)

Abstract

Aims: Markerless motion analysis (MMA) systems are being used extensively in the area of sports medicine and physiotherapy. The purpose of this study was to compare leg length measurements (LLM) and varus/valgus knee measurements (VVM) performed clinically, radiologically and using MMA in patients being assessed for potential total knee arthroplasty (TKR).between mean LLM calculated clinically, radiologically and using MMA (all p < 0.05). Discussion & conclusion: Discrepanices exist in LLM and VVM when evaluated using clinical, radiological and MMA modalities. Therefore, this study suggests that MMA alone may not be a suitable modality for assessment of patients for TKR, with a combination of two or more evaluation modalities recommended at present. Level of evidence: IV Case Series.

Original languageEnglish
Pages (from-to)45-48
Number of pages4
JournalJournal of Orthopaedics
Volume26
DOIs
Publication statusPublished - 1 Jul 2021

Keywords

  • Markerless motion analysis
  • TKR
  • Total knee replacement