Skip to main navigation Skip to search Skip to main content

Analysis of CyberKnife intracranial treatment plans using ICRU 91 dose reporting: A retrospective study

  • University of Galway
  • Galway Clinic
  • Blackrock Health Hermitage Medical Clinic

Research output: Contribution to a Journal (Peer & Non Peer) › Article › peer-review

2 Citations (Scopus)

Abstract

ICRU 91, published in 2017, is an international standard for prescribing, recording, and reporting stereotactic treatments. Since its release, there has been limited research published on the implementation and impact of ICRU 91 on clinical practice. This work provides an assessment of the recommended ICRU 91 dose reporting metrics for their use in clinical treatment planning. A set of 180 intracranial stereotactic treatment plans for patients treated by the CyberKnife (CK) system were analyzed retrospectively using the ICRU 91 reporting metrics. The 180 plans comprised 60 trigeminal neuralgia (TGN), 60 meningioma (MEN), and 60 acoustic neuroma (AN) cases. The reporting metrics included the planning target volume (PTV) near-minimum dose ((Formula presented.)), near-maximum dose ((Formula presented.)), and median dose ((Formula presented.)), as well as the gradient index (GI) and conformity index (CI). The metrics were assessed for statistical correlation with several treatment plan parameters. In the TGN plan group, owing to the small targets, (Formula presented.) was greater than (Formula presented.) in 42 plans, whereas both metrics were not applicable in 17 plans. The (Formula presented.) metric was predominantly influenced by the prescription isodose line (PIDL). The GI was significantly dependent on target volume in all analyses performed, where the variables were inversely related. The CI was only dependent on target volume in treatment plans for small targets. The ICRU 91 (Formula presented.) and (Formula presented.) metrics breakdown in plans for small target volumes below 1 cm3; the Min and Max pixel should be reported in such cases. The (Formula presented.) metric is of limited use for treatment planning. Given their volume dependence, the GI and CI metrics could potentially serve as plan evaluation tools in the planning of the sites analyzed in this study, which would ultimately improve treatment plan quality.

Original languageEnglish
Article numbere13932
JournalJournal of Applied Clinical Medical Physics
Volume24
Issue number6
DOIs
Publication statusPublished - Jun 2023

Keywords

  • CyberKnife
  • ICRU 91
  • conformity index
  • gradient index
  • stereotactic
  • treatment planning