Skip to main navigation Skip to search Skip to main content

Adaptive optics for ophthalmic imaging

  • Luis Diaz-Santana
  • , Paul Gasson
  • , Ian Munro
  • , Chris Dainty
  • City University London
  • Imperial College London

Research output: Chapter in Book or Conference Publication/ProceedingConference Publicationpeer-review

Abstract

In this report we present an adaptive optics system with a closed-loop bandwidth between 12 Hz and 25 Hz. This system scans the wavefront sensing beam at 1 kHz over a maximum angle of 1 degree and de-scans it on the exit path to eliminate speckle, as previously done by Hofer et.al to study the dynamics of ocular aberrations without correction. This has improved the signal to noise ratio in the wavefront sensor signal, hence ensuring a reduction of ocular aberrations in all of the subjects tested. The system uses a Shack-Hartmann wavefront sensor and a DALSA CCD camera. MTF was reconstructed from the images using a Wiener de-convolution algorithm assuming a quasidiffraction limited first pass, followed by a discrete Fourier transform.

Original languageEnglish
Title of host publication2003 Conference on Lasers and Electro-Optics Europe, CLEO/EUROPE 2003
Pages723
Number of pages1
DOIs
Publication statusPublished - 2003
Event2003 Conference on Lasers and Electro-Optics Europe, CLEO/EUROPE 2003 - Munich, Germany
Duration: 22 Jun 200327 Jun 2003

Publication series

NameConference on Lasers and Electro-Optics Europe - Technical Digest

Conference

Conference2003 Conference on Lasers and Electro-Optics Europe, CLEO/EUROPE 2003
Country/TerritoryGermany
CityMunich
Period22/06/0327/06/03

Fingerprint

Dive into the research topics of 'Adaptive optics for ophthalmic imaging'. Together they form a unique fingerprint.

Cite this