A compact violet diode laser based fluorescence lifetime microscope

  • Alan Ryder
  • , Thomas J. Glynn
  • , Milosz A. Przyjalgowski
  • , Boguslaw Szczupak

Research output: Contribution to conference (Published)International Refereed Conference Paperpeer-review

20 Citations (Scopus)

Abstract

We have constructed a time-correlated single-photon counting (TCSPC) microscope system using a pulsed violet-laser diode for measuring fluorescence lifetimes below 1 ns. These compact; cool, and fast-pulsed laser diodes are much more suitable for lifetime measurements than the traditional mode-locked lasers used in the past. The laser source (LDH-400, PicoQuant GmbH) is mounted on an optical table above the microscope, and the laser beam is directed into an Olympus BX-60 microscope by means of a dichroic beamsplitter set at 45 degrees. The system operates in backscattering mode with the fluorescence emission passing back through the dichroic beamsplitter and focused into a 100-monochromator with a PMT detector. This prototype instrument is compact (80 70 70 cm) and is nearly fully computer controlled by means of a SPC-730 (Becker amp; Hickl) PC card. We include preliminary results showing the instrument response function (IRF) of the system, and some of the factors have been adjusted to minimize the temporal width of the IRF. The instrument has been validated using a series of standard fluorophores at different emission wavelengths.
Original languageEnglish (Ireland)
Pages177-180
Number of pages4
DOIs
Publication statusPublished - 1 Jun 2002

Keywords

  • Fluorescence
  • Laser diode
  • Lifetime
  • Microscopy

Authors (Note for portal: view the doc link for the full list of authors)

  • Authors
  • Ryder,A.G, Glynn,T.J, Przyjalgowski, M, Szczupak,B.

Fingerprint

Dive into the research topics of 'A compact violet diode laser based fluorescence lifetime microscope'. Together they form a unique fingerprint.

Cite this